Title of article :
Flow-injection in-line complexation for ion-pair reversed phase high performance liquid chromatography of some metal-4-(2-pyridylazo) resorcinol chelates
Author/Authors :
Srijaranai، نويسنده , , Supalax and Chanpaka، نويسنده , , Saiphon and Kukusamude، نويسنده , , Chutima and Grudpan، نويسنده , , Kate، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2006
Abstract :
Flow injection (FI) was coupled to ion-pair reversed phase high performance liquid chromatography (IP-RPHPLC) for the simultaneous analysis of some metal-4-(2-pyridylazo) resorcinol (PAR) chelates. A simple reverse flow injection (rFI) set-up was used for in-line complexation of metal-PAR chelates prior to their separation by IP-RPHPLC. The rFI conditions were: injection volume of PAR 85 μL, flow rate of metal stream 4.5 mL min−1, concentration of PAR 1.8 × 10−4 mol L−1 and the mixing coil length of 150 cm. IP-RPHPLC was carried out using a C18 μBondapak column with the mobile phase containing 37% acetonitrile, 3.0 mmol L−1 acetate buffer pH 6.0 and 6.2 mmol L−1 tetrabutylammonium bromide (TBABr) at a flow rate of 1.0 mL min−1 and visible detection at 530 and 440 nm. The analysis cycle including in-line complexation and separation by IP-RPHPLC was 16 min, which able to separate Cr(VI) and the PAR chelates of Co(II), Ni(II) and Cu(II).
Keywords :
Flow Injection , Ion pair reversed phase high performance liquid chromatography , Metal-PAR chelates , In-line complexation