Periodically interrupted amperometry at membrane coated electrodes: A simplified pulsed amperometry
Author/Authors :
Takلtsy، نويسنده , , Anikَ and Csَka، نويسنده , , Balلzs and Nagy، نويسنده , , Lيvia and Nagy، نويسنده , , Géza، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2006
Abstract :
Amperometric detection combined with separation technique or with selective molecular recognition step can be very effective solving quantitative analytical tasks. When the amperometric working electrode surface needs cleaning or reactivation, pulsed amperometric technique can be the choice. Coating working electrodes with different sensitizing or protecting layer is quite common in the practice of voltammetric analysis. In these studies the behavior of coated electrodes using a simplified pulsed amperometric working program which can be named periodically interrupted amperometric (PIA) detection has been investigated. Rotating platinum, and carbon paste electrodes coated with dialysis film or porcine intestinal membrane were used in the experiments. The signal in case of electrochemical oxidation of hydrogen peroxide and ascorbic acid at convective conditions has been evaluated. The signal, obtained with conventional amperometry has been compared with signal collected with a periodically interrupted amperometric measuring program, allowing time for the diffusion to reload the diffusion layer at the electrode surface. The sensitivity and the lower limit of detection (4.5 × 10−7 M for ascorbic acid and 2 × 10−6 M for H2O2) proved superior in case of the periodically interrupted amperometry.
Keywords :
ascorbic acid , Hydrogen peroxide , Pulsed amperometry , Sensitivity , Lower detection limit