Title of article :
Conductometric nitrate biosensor based on methyl viologen/Nafion®/nitrate reductase interdigitated electrodes
Author/Authors :
Xuejiang، نويسنده , , Wang and Dzyadevych، نويسنده , , Sergei V and Chovelon، نويسنده , , Jean-Marc and Renault، نويسنده , , Nicole Jaffrezic and Ling، نويسنده , , Chen and Siqing، نويسنده , , Xia and Jianfu، نويسنده , , Zhao، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2006
Pages :
6
From page :
450
To page :
455
Abstract :
A highly sensitive, fast and stable conductometric enzyme biosensor for determination of nitrate in water is reported for the first time. The biosensor electrodes were modified by methyl viologen mediator mixed with nitrate reductase (NR) from Aspergillus niger by cross-linking with glutaraldehyde in the presence of bovine serum albumin and Nafion® cation-exchange polymer. The process parameters for the fabrication of the enzyme electrode and various experimental variables such as pH, the enzyme loading and time of immobilization in glutaralaldehyde vapor were investigated with regard to their influence on sensitivity, limit of detection, dynamic range and operational and storage stability. The biosensor can reach 95% of steady-state conductance value in about 15 s. Linear calibration in the range of 0.02 and 0.25 mM with detection limits of 0.005 mM nitrate was obtained with a signal-to-noise ratio of 3. When stored in 5 mM phosphate buffer (pH 7.5) at 4 °C, the sensor showed good stability over 2 weeks.
Keywords :
Methyl viologen , nitrate , interdigitated electrodes , Conductimetric biosensor , nitrate reductase
Journal title :
Talanta
Serial Year :
2006
Journal title :
Talanta
Record number :
1649950
Link To Document :
بازگشت