• Title of article

    Determination of Se(IV) by anodic stripping voltammetry using gold electrodes made from recordable CDs

  • Author/Authors

    Pereira، نويسنده , , Claudete Fernandes and Gonzaga، نويسنده , , Fabiano Barbieri and Guaritل-Santos، نويسنده , , Antonio Moraes and SouzaDe، نويسنده , , Jurandir Rodrigues SouzaDe، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2006
  • Pages
    5
  • From page
    877
  • To page
    881
  • Abstract
    This work presents a simple, reproducible and low cost method, employing differential pulse anodic stripping voltammetry, for determination of selenium(IV). A gold electrode obtained from recordable compact disks (CD-R) was used to evaluate the voltammetric behavior of the metallic ion in 0.1 mol L−1 HClO4. To evaluate the voltammetric behavior of Se(IV), parameters such as deposition potential and deposition time were optimized. A wide linear response range, from 0.5 to 291 ng mL−1, was obtained using a 5.0 mm diameter gold electrode. Recovery tests for Se(IV) utilizing standard reference solutions provided values between 94 and 96%.
  • Keywords
    Recordable CDs , Selenium , Gold electrode
  • Journal title
    Talanta
  • Serial Year
    2006
  • Journal title
    Talanta
  • Record number

    1650156