Title of article :
In-channel indirect amperometric detection of heavy metal ions for electrophoresis on a poly(dimethylsiloxane) microchip
Author/Authors :
Li، نويسنده , , Xin-Ai and Zhou، نويسنده , , Dongmei and Xu، نويسنده , , Jing-Juan and Chen، نويسنده , , Hong-Yuan، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2007
Pages :
6
From page :
1130
To page :
1135
Abstract :
In-channel indirect amperometric detection mode for microchip capillary electrophoresis with positive separation electric field is successfully applied to some heavy metal ions. The influences of separation voltage, detection potential, the concentration and pH value of running buffer on the response of the detector have been investigated. An optimized condition of 1200 V separation voltage, −0.1 V detection potential, 20 mM (pH 4.46) running buffer of 2-(N-morpholino)ethanesulfonic acid (MES) + l-histidine (l-His) was selected. The results clearly showed that Pb2+, Cd2+, and Cu2+ were efficiently separated within 80 s in a 3.7 cm long native separation PDMS/PDMS channel and successfully detected at a single carbon fibre electrode. The theoretical plate numbers of Pb2+, Cd2+, and Cu2+ were 1.2 × 105, 2.5 × 105, and 1.9 × 105 m−1, respectively. The detection limits for Pb2+, Cd2+, and Cu2+ were 1.3, 3.3 and 7.4 μM (S/N = 3).
Keywords :
Microchip capillary electrophoresis , Poly(dimethylsiloxane) , Carbon fibre electrode , Heavy metal ions , Indirect amperometric detection mode
Journal title :
Talanta
Serial Year :
2007
Journal title :
Talanta
Record number :
1651486
Link To Document :
بازگشت