Title of article
Voltammetric determination of hafnium with calconcarboxylic acid
Author/Authors
Liliya and Levytska، نويسنده , , Halyna and Orshuliak، نويسنده , , Oksana، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2007
Pages
3
From page
1441
To page
1443
Abstract
The interaction of Hf(IV) with calconcarboxylic acid (1-(2-hydroxy-4-sulfo-1-naphtylazo)-3-naphtolic acid) was investigated by cyclic voltammetry varying various factors (pH, ionic strength, contents of ethanol and scan rate). Optimal conditions of Hf(IV) determination in the presence of calconcarboxylic acid were found: acetic buffer solution pH 2.6, scan rate 0.5 Vs−1. The detection limit of Hf(IV) concentration was 2.46 × 10−7 mol L−1. The influence of foreign ions on the Hf(IV) determination was studied. It was established that some metals like cadmium, nickel, zinc, copper and titanium could be determined with Hf(IV) simultaneously. The method of voltammetric determination of hafnium was tested on model solutions and used in the determination of Hf(IV) in terbium-base alloy.
Keywords
Cyclic voltammetry , Calconcarboxylic acid (CCA) , hafnium
Journal title
Talanta
Serial Year
2007
Journal title
Talanta
Record number
1651629
Link To Document