Title of article :
New automatized method with amperometric detection for the determination of azithromycin
Author/Authors :
Palomeque، نويسنده , , Miriam E. and Ortيz، نويسنده , , Patricia I.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2007
Pages :
5
From page :
101
To page :
105
Abstract :
A FIA–amperometric method for azithromycin determination was developed. A working glassy carbon electrode and a Ag/AgCl/NaCl (3 M) reference electrode were used. The determination is based on the electrochemical oxidation of the azithromycin at 0.9 V in Britton–Robinson buffer solution (pH 8.0). Due to the adsorption of the reaction products on the electrode surface, an effective cleaner cycle was implemented. By using the optimum chemical and FIA conditions, a concentration linear range of 1.0–10.0 mg L−1 and a detection limit (LOD) of 0.76 mg L−1 are obtained. The method was validated and satisfactorily applied to the determination of azithromycin in pharmaceutical formulations.
Keywords :
azithromycin , FIA , Pharmaceutical Preparations , Amperometry
Journal title :
Talanta
Serial Year :
2007
Journal title :
Talanta
Record number :
1652003
Link To Document :
بازگشت