• Title of article

    New automatized method with amperometric detection for the determination of azithromycin

  • Author/Authors

    Palomeque، نويسنده , , Miriam E. and Ortيz، نويسنده , , Patricia I.، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2007
  • Pages
    5
  • From page
    101
  • To page
    105
  • Abstract
    A FIA–amperometric method for azithromycin determination was developed. A working glassy carbon electrode and a Ag/AgCl/NaCl (3 M) reference electrode were used. The determination is based on the electrochemical oxidation of the azithromycin at 0.9 V in Britton–Robinson buffer solution (pH 8.0). Due to the adsorption of the reaction products on the electrode surface, an effective cleaner cycle was implemented. By using the optimum chemical and FIA conditions, a concentration linear range of 1.0–10.0 mg L−1 and a detection limit (LOD) of 0.76 mg L−1 are obtained. The method was validated and satisfactorily applied to the determination of azithromycin in pharmaceutical formulations.
  • Keywords
    azithromycin , FIA , Pharmaceutical Preparations , Amperometry
  • Journal title
    Talanta
  • Serial Year
    2007
  • Journal title
    Talanta
  • Record number