Flow injection analysis of ethyl xanthate by in-line dialysis and UV spectrophotometric detection
Author/Authors
Fontenele، نويسنده , , Rinaldo S. and Hidalgo، نويسنده , , Pilar and Gutz، نويسنده , , Ivano G.R. and Pedrotti، نويسنده , , Jairo J.، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2007
Pages
6
From page
1017
To page
1022
Abstract
A simple and sensitive spectrophotometric flow method for determination of low concentrations of the flotation collector O-ethyldithiocarbonate (ethyl xanthate, CH3CH2-O-CS2−) in solutions is described. The method is based on ethyl xanthate detection at 301 nm in medium of NaOH 50 mmol L−1. By injection of 200 μL of sample, the analytical method shows linear response for the ethyl xanthate concentration from 0.5 up to 500 μmol L−1. Successive injections of 4 μmol L−1 ethyl xanthate (n = 23) show a coefficient of variation lower than 0.6%, denoting high repeatability. The detection limit is 0.3 μmol L−1. At a flow rate of 2.0 mL min−1, a frequency of 120 injections/h of ethyl xanthate can be attained. By introduction of a tangential dialysis cell in the FIA system, the manual sample filtration step with 0.22 μm filter was eliminated and the residual interference of suspended material, was completely overcome even for unfiltered sludge suspension samples, an important advantage that compensates for the frequency reduction to 25 injections/h elevation and detection limit elevation to 2 μmol L−1, still outreaching for many applications. Potential applications of the method embrace the at line determination of ethyl xanthate in the ore processing industry, control of the concentration at its optimal level during the flotation process, as well as monitoring of residues in the effluents.
Keywords
Spectrophotometry , Adsorption , Dialysis , Pyrite , Ethyl xanthate , Flow Analysis
Journal title
Talanta
Serial Year
2007
Journal title
Talanta
Record number
1652415
Link To Document