• Title of article

    Aluminum(III) selective potentiometric sensor based on morin in poly(vinyl chloride) matrix

  • Author/Authors

    Gupta، نويسنده , , Vinod K. and Jain، نويسنده , , Ajay K. and Maheshwari، نويسنده , , Gaurav، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2007
  • Pages
    5
  • From page
    1469
  • To page
    1473
  • Abstract
    Al3+ selective sensor has been fabricated from poly(vinyl chloride) (PVC) matrix membranes containing neutral carrier morin as ionophore. Best performance was exhibited by the membrane having composition as morin:PVC:sodium tetraphenyl borate:tri-n-butylphosphate in the ratio 5:150:5:150 (w/w, mg). This membrane worked well over a wide activity range of 5.0 × 10−7 to 1.0 × 10−1 M of Al3+ with a Nernstian slope of 19.7 ± 0.1 mV/decade of Al3+ activity and a limit of detection 3.2 × 10−7 M. The response time of the sensor is ∼5 s and membrane could be used over a period of 2 months with good reproducibility. The proposed sensor works well over a pH range (3.5–5.0) and demonstrates good discriminating power over a number of mono-, di- and trivalent cations. The sensor can also be used in partially non-aqueous media having up to 20% (v/v) methanol, ethanol or acetone content with no significant change in the value of slope or working activity range. The sensor has also been used in the potentiometric titration of Al3+ with EDTA and for its determination in zinc plating mud and red mud.
  • Keywords
    Nernstian slope , Aluminum selective sensor , morin , Poly(vinyl chloride) (PVC) , ionophore
  • Journal title
    Talanta
  • Serial Year
    2007
  • Journal title
    Talanta
  • Record number

    1652615