Title of article :
Instrumental neutron activation analysis of mass fractions of toxic metals in plastic
Author/Authors :
Park، نويسنده , , Kwangwon and Kang، نويسنده , , Namgoo، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2007
Pages :
4
From page :
791
To page :
794
Abstract :
It is very challenging to decompose a plastic product for the purpose of analysis of hazardous elements contained. To circumvent such technical problem, it is imperative that an analyst employ a nondestructive analytical method free of any pretreatments. The analytical results of the concentrations of toxic metals such as Cd and Cr in polypropylene for seven samples at two different levels were obtained using the instrumental neutron activation analysis. This work was intended ultimately to establish certified reference materials (CRMs) of these metals in the polypropylene, traceable to the SI. The uncertainties associated with the analytical procedures were estimated in accordance with the ISO guideline. The results were subsequently validated by a comparison with those for CRM-680 and −681 of the Bureau Communautaire de Reference (BCR), which demonstrated acceptable agreement within their uncertainty ranges.
Keywords :
uncertainty , Polypropylene , NEUTRON ACTIVATION ANALYSIS , certified reference material , Hazardous element
Journal title :
Talanta
Serial Year :
2007
Journal title :
Talanta
Record number :
1653148
Link To Document :
بازگشت