• Title of article

    Multi-elemental analysis of produced water by synchrotron radiation total reflection X-ray fluorescence

  • Author/Authors

    Pimentel، نويسنده , , P.M. and Anjos، نويسنده , , M.J. and Melo، نويسنده , , D.M.A. and Melo، نويسنده , , M.A.F. and Gonçalves Jr.، نويسنده , , L.M. and Silva Jr.، نويسنده , , C.N. and Lopes، نويسنده , , R.T.، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2008
  • Pages
    5
  • From page
    1231
  • To page
    1235
  • Abstract
    The synchrotron radiation total reflection X-ray fluorescence (SRTXRF) technique was used for the analysis of heavy metals in produced water samples from oil field in Rio Grande do Norte, in order to determine potential sources of pollution. Since the inorganic components in produced water generally resembling sea water, pre-concentration procedures have been applied to increase the concentration of the analyte of interest and to minimize the salt matrix effects. This technique allows us to determine the contents of V, Cr, Mn, Fe, Co, Ni, Cu, Zn, Hg and Pb in 20 produced water samples. The great majority of the sampling points presented low elemental concentration value. However, in some sample, the Fe, Ni, Cu, Zn and Hg concentration were higher than the established limits by the Brazilian legislation.
  • Keywords
    SRTXRF , Heavy metals , Produced water
  • Journal title
    Talanta
  • Serial Year
    2008
  • Journal title
    Talanta
  • Record number