• Title of article

    Parameters affecting the determination of mercury by anodic stripping voltammetry using a gold electrode

  • Author/Authors

    Giacomino، نويسنده , , Agnese and Abollino، نويسنده , , Ornella and Malandrino، نويسنده , , Mery and Mentasti، نويسنده , , Edoardo، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2008
  • Pages
    8
  • From page
    266
  • To page
    273
  • Abstract
    The electrochemical determination of aqueous Hg(II) by anodic stripping voltammetry (ASV) at a solid gold electrode is described. The aim of this work is to optimise all factors that can influence this determination. Potential wave forms (linear sweep, differential pulse, square wave), potential scan parameters, deposition time, deposition potential and surface cleaning procedures were examined for their effect on the mercury peak shape and intensity. Five supporting electrolytes were tested. The best responses were obtained with square wave potential wave form and diluted HCl as supporting electrolyte. Electrochemical and mechanical surface cleaning, aimed at removing the amount of mercury deposited onto the gold surface, were necessary for obtaining a good performance of the electrode. Response linearity, repeatability, accuracy and detection limit were also evaluated.
  • Keywords
    mercury , Anodic stripping voltammetry , Gold electrode , Supporting electrolyte , Wave form
  • Journal title
    Talanta
  • Serial Year
    2008
  • Journal title
    Talanta
  • Record number