Title of article :
Adjusting mobility scales of ion mobility spectrometers using 2,6-DtBP as a reference compound
Author/Authors :
Viitanen، نويسنده , , A.-K. and Mauriala، نويسنده , , T. Yli-Mattila ، نويسنده , , T. and Adamov، نويسنده , , A. and Pedersen، نويسنده , , C.S. and Mنkelن، نويسنده , , J.M. and Marjamنki، نويسنده , , M. and Sysoev، نويسنده , , A. and Keskinen، نويسنده , , J. S. Kotiaho، نويسنده , , T.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2008
Abstract :
Performance of several time-of-flight (TOF) type ion mobility spectrometers (IMS) was compared in a joint measurement campaign and their mobility scales were adjusted based on the measurements. A standard reference compound 2,6-di-tert butylpyridine (2,6-DtBP) was used to create a single peak ion mobility distribution with a known mobility value. The effective length of the drift tube of each device, considered here as an instrument constant, was determined based on the measurements. Sequentially, two multi-peaked test compounds, DMMP and DIMP, were used to verify the performance of the adjustment procedure in a wider mobility scale. By determining the effective drift tube lengths using 2,6-DtBP, agreement between the devices was achieved. The determination of effective drift tube lengths according to standard reference compound was found to be a good method for instrument inter-comparison. The comparison procedure, its benefits and shortcomings as well as dependency on accuracy of literature value are discussed along with the results.
Keywords :
Instrument comparison , Time-of-flight , 6-DtBP , Reference compound , 2 , IMS