Title of article :
Determination of Hg2+ on poly(vinylferrocenium) (PVF+)-modified platinum electrode
Author/Authors :
Mutlu Sonmez Celebi، نويسنده , , Mutlu and ?zy?rük، نويسنده , , Haluk and Y?ld?z، نويسنده , , Attila and Abac?، نويسنده , , Serdar، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2009
Pages :
5
From page :
405
To page :
409
Abstract :
A new surface based on poly(vinylferrocenium) (PVF+)-modified platinum electrode was developed for determination of Hg2+ ions in aqueous solutions. The polymer was electrodeposited on platinum electrode by constant potential electrolysis as PVF+ClO4−. Cl− ions were then attached to the polymer matrix by anion exchange and the modified electrode was dipped into Hg2+ solution. Hg2+ was preconcentrated at the polymer matrix by adsorption and also complexation reaction with Cl−. Detection of Hg2+ was carried out by differential pulse anodic stripping voltammetry (DPASV) after reduction of Hg2+. Mercury ions as low as 5 × 10−10 M could be detected with the prepared electrode and the relative standard deviation was calculated as 6.35% at 1 × 10−6 M concentration (n = 6). Interferences of Ag+, Pb2+ and Fe3+ ions were also studied at two different concentration ratios with respect to Hg2+. The developed electrode was applied to the determination of Hg2+ in water samples.
Keywords :
Mercury determination , PVF+ , Anodic stripping , Interference , aqueous solution , complexation
Journal title :
Talanta
Serial Year :
2009
Journal title :
Talanta
Record number :
1656844
Link To Document :
بازگشت