Title of article :
Capillary electrophoresis-mass spectrometry using an in-line sol–gel concentrator for the determination of methionine enkephalin in cerebrospinal fluid
Author/Authors :
Ramautar، نويسنده , , R. and Ratnayake، نويسنده , , C.K. and Somsen، نويسنده , , G.W and de Jong، نويسنده , , G.J.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2009
Abstract :
In this study, a CE-MS method using a monolithic sol–gel concentrator for in-line solid-phase extraction (SPE) is evaluated for the analysis of methionine enkephalin in biological samples. Operational SPE parameters such as sample pH, loading volume, elution volume and composition have been studied. After optimization of the in-line preconcentration methodology, a 40-fold preconcentration was demonstrated for a methionine enkephalin test solution using a loading volume of 3200 nL. The method was linear in the range from 62.5 to 1000 ng/mL (R2 > 0.99). R.S.D. values for migration times and peak areas were 1.2% and 8.4%, respectively. Finally, the analysis of cerebrospinal fluid samples spiked with methionine enkephalin and deproteinized with perchloric acid (1:1, v/v) showed a detection limit (S/N = 3) of approximately 1 ng/mL (ca. 5 nM). The recoveries of methionine enkephalin for three concentration levels (100, 10 and 1 ng/mL) were in the range of 74–91%, demonstrating the promising potential of the methodology for the analysis of biological samples.
Keywords :
cerebrospinal fluid , Preconcentration , Methionine enkephalin , Capillary electrophoresis-mass spectrometry , Sol–gel concentrator