Title of article :
The petentiometric determination of peroxide hydrogen and glucose on the glassy electrode modified by the calix[4]arene
Author/Authors :
Jin، نويسنده , , Gendi and Du، نويسنده , , Shi and Hu، نويسنده , , Xiaoya، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2009
Abstract :
A new petentiometric method to determine peroxide hydrogen and glucose had been studied. This method had been applied on the petentiometric determination of peroxide hydrogen and glucose in the total ionic strength adjustment buffer (TISAB) (pH 7.5) solution with the glassy electrode modified by the calix[4]arene. The glassy carbon electrode covered with the calix[4]arene depended on the H2O2 concentration in the range of log[H2O2] from −3.3 to −1.2 in the solution of TISAB (pH 7.5) with nearly Nernstian slope of about 65.6 ± 3 mV and the detection limit of peroxide hydrogen was 4.0 × 10−5 mol L−1. The glassy carbon electrode covered with the calix[4]arene depended on the glucose concentration in the range of log[glucose] from −3.6 to −2.8 in the solution of TISAB (pH 7.5) with nearly Nernstian slope of about 50.2 ± 2 mV and the detection limit of glucose was 2.0 × 10−5 mol L−1. The electrode had the good selectivity, sensitivity, stability and repeatability.
Keywords :
Peroxide Hydrogen , Glassy carbon electrode , Glucose , Petentiometric determination