Title of article :
Process sampling module coupled with purge and trap–GC–FID for in situ auto-monitoring of volatile organic compounds in wastewater
Author/Authors :
Liu، نويسنده , , Hsin-Wang and Liu، نويسنده , , Yung-Tsun and Wu، نويسنده , , Bei-Zen and Nian، نويسنده , , Hung-Chi and Chen، نويسنده , , Hsing-Jung and Chiu، نويسنده , , Kong-Hwa and Lo، نويسنده , , Jiunn-Guang Lo، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2009
Pages :
6
From page :
903
To page :
908
Abstract :
An automatic sampling device, i.e., process sampling module (PSM), connected with a purge and trap–GC–FID system has been developed for real-time monitoring of VOCs in wastewater. The system was designed to simultaneously monitor 17 compounds, including one polar compound, i.e., acetone, and 16 non-polar compounds. The trapping tube is packed with two adsorbents, Carbopack B and Carbosieve III, to trap target compounds. For the purpose of in situ monitoring, the flush valve of the sampling tube is composed of two two-way valves and a time controller to prevent absorption interference of the residue. The optimal conditions for the analytical system include a 12 min purge time at a temperature of 60 °C, and 4 min of desorption time with a desorption temperature of 260 °C. Good chromatograms have been obtained with the analytical system even if a cryogenic device and de-misting were not used. The relative standards deviation (RSD) of the system is between 2% and 13.4%, and accuracies between 0.3 and 23.5% have been achieved. The detection limits of the method range from 0.32 to 2.39 ppb. In this system, the four parts, i.e., PSM, P&T, GC, and FID, were simple, reliable and rugged. Also, the interface of these four parts was simple and dependable.
Keywords :
P& , Process sampling module , T–GC–FID , VOCS , In situ auto-monitoring , water
Journal title :
Talanta
Serial Year :
2009
Journal title :
Talanta
Record number :
1659126
Link To Document :
بازگشت