Title of article
Preparation of a new solid state fluoride ion selective electrode and application
Author/Authors
Somer، نويسنده , , Güler and Kalayc?، نويسنده , , ?ükrü and Ba?ak، نويسنده , , ?brahim، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2010
Pages
4
From page
1129
To page
1132
Abstract
A new solid state fluoride ion selective electrode composed of 70% Ag2S, 10% Cu2S and 20% CaF2 has been developed. An analytically useful potential change occurred, from 1 × 10−6 to 1 × 10−1 M fluoride ion. The slope of the linear portion (1 × 10−1–1 × 10−5 M) was about 26 ± 2 mV/10-fold concentration changes in fluoride. It was found that pH change between 1 and 8 had no effect on the potential of the electrode. There was no interference of most common cations such as K+, Na+, Ca2+ and Mg2+ and anions such as Cl−, NO3−, SO42− and PO43−. The lifetime of the electrode was more than 2 years, when used at least 4–5 times a day, and the response time was about 60 s.
asurements were made at constant ionic strength (0.1 M NaNO3) and at room temperature. This electrode has been used for the determination of fluoride ion in Ankara city tap water and in bottled spring water using standard addition method. The validation of the electrode has been made with a commercial fluoride ion selective electrode (Orion) and high consistency was obtained.
Keywords
Fluoride electrode , Fluoride in water , solid state , Determination of fluoride
Journal title
Talanta
Serial Year
2010
Journal title
Talanta
Record number
1659236
Link To Document