Title of article :
Advantages of using a modified orthogonal sampling configuration originally designed for LC–ESI-MS to couple CE and MS for the determination of antioxidant phenolic compounds found in virgin olive oil
Author/Authors :
Nevado، نويسنده , , Juan José Berzas and Peٌalvo، نويسنده , , Gregorio Castaٌeda and Robledo، نويسنده , , Virginia Rodrيguez and Martيnez، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2010
Pages :
7
From page :
548
To page :
554
Abstract :
A ThermoFinnigan sheath liquid flow capillary electrophoresis–mass spectrometry system designed for coupling via a co-axial interface was coupled through an adapted via an alternative, commercially available interface for orthogonal sampling. The affordable, reversible structural alterations made in the commercial LC–MS interface resulted in improved analytical performance. sults of a conventional capillary electrophoresis (CE) method using a commercial co-axial source to determine antioxidant phenolic acids present in virgin olive oil, were compared with those obtained by using a modified orthogonal sampling position. In both cases, separations were done using a 10 mM ammonium acetate/ammonium hydroxide buffer solution at pH 10.0 and a constant applied voltage of 25 kV. The operating variables for the mass spectrometry interface were re-optimized for the modified orthogonal orientation. This allowed the sheath liquid, sheath gas flow rates and capillary voltage to be lowered with respect to the co-axial coupling configuration. In addition, the orthogonal sampling position provided a higher selectivity by effect of ion sampling excluding larger droplets—with an increased momentum along the axis—which were drained through the sink at the bottom of the ion source. Also, the new configuration facilitated sample ionization, improved electrospray stability and led to stronger signals as a result. w system was validated in terms of precision (repeatability), linearity, and limits of detection and quantification. A comparison of the validation data with the results previously obtained by using a commercial co-axial configuration revealed the adapted orthogonal sampling position to provide better repeatability in both migration times and relative peak areas (<1% and 7% respectively with n = 15 replicates), a good linear range (with levels in the microgram-per-litre region) and lower limits of detection—especially for the compounds detected with the lowest sensitivity when co-axial ESI was used, as HFA, GEN, FER and VAN finding LOD among 24–3.0 μg L−1 respectively.
Keywords :
antioxidants , Orthogonal coupling , Electrospray source , Phenolic acids , CE–ESI-MS
Journal title :
Talanta
Serial Year :
2010
Journal title :
Talanta
Record number :
1660232
Link To Document :
بازگشت