Title of article :
An in-tube SPME device for the selective determination of chlorophyll a in aquatic systems
Author/Authors :
Vitta، نويسنده , , Y. and Moliner-Martيnez، نويسنده , , Y. and Campيns-Falcَ، نويسنده , , P. and Cuervo، نويسنده , , A. Fernلndez، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2010
Abstract :
We report a new device for the estimation of the content of chlorophyll a pigment in water samples as an indicator of water quality. The extraction of the pigment from water was also optimized. 10 mL of water was filtered through a nylon filter (45 μm pore size and 13 mm of diameter), after the chlorophylls were dissolved by immersing the filter in 1 mL of a low non-hazardous solvent as ethanol. An in-valve in-tube SPME device coupled to capillary liquid chromatography with diode array detection was designed. A capillary column of 70 cm in length (0.32 mm i.d. coated with 5% diphenyl–95% polydimethylsiloxane, 3 μm coating thickness) was used as the loop of the injection valve for preconcentration and a Zorbax SB C18 (SiO2-based) 150 mm × 0.5 mm i.d., 5 μm column (Agilent) was used as analytical column. The achieved detection limit was 0.05 μg L−1 and the working range of concentrations was 0.1–1 μg L−1. % RSD values between 2 and 11 were obtained. Chlorophyll a in several water matrices was determined with good results in presence of other pigments such as chlorophyll b, pheophytin a and pheophytin b.
Keywords :
In-tube solid phase microextraction (IT-SPME) , Water quality , Chlorophyll a , Capillary LC