Title of article :
Kinetic catalytic determination of trace levels of iodide based on the oxidation of basic dyes with hydrogen peroxide monitored potentiometrically using simple PVC electrodes
Author/Authors :
Khaled، نويسنده , , Elmorsy and El-Ries، نويسنده , , M.A. and Zidane، نويسنده , , F.I. and Ibrahim، نويسنده , , S.A. and Abd-Elmonem، نويسنده , , M.S.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2011
Pages :
6
From page :
1538
To page :
1543
Abstract :
Four sensitive catalytic potentiometric methods have been developed for trace levels determination of iodide based on its catalytic effects on the oxidation of four dyes: viz. variamine blue (VB), rhodamine B (RB), methylene blue (MB) and malachite green (MG), with H2O2 in H3PO4 medium at 25 ± 0.5 °C. The catalyzed reaction rates were estimated potentiometrically by monitoring the potential of the corresponding dye-PVC ion selective electrodes. To select the optimized reaction conditions offering the highest sensitivity of the method, parallel studies were carried out on each dye catalyzed reaction including: the effect of reactant concentration, reaction medium and temperature. The working calibration curves were linear over the concentration range from 0.32 to 2.54 mg L−1 iodide for VB method and from 3.2 to 12.7 mg L−1 for other ones. The tolerance limits of more than 20 interfering species were listed indicating the high selectivity of the method. Trace iodide in edible salt and pharmaceutical samples was determined without the need for separation or preconcentration procedures.
Keywords :
PVC electrodes , Variamine blue (VB) , Edible salt , Catalytic potentiometry , iodide , Pharmaceutical Preparations
Journal title :
Talanta
Serial Year :
2011
Journal title :
Talanta
Record number :
1661669
Link To Document :
بازگشت