Title of article
Fast and direct determination of butylated hydroxyanisole in biodiesel by batch injection analysis with amperometric detection
Author/Authors
Tormin، نويسنده , , Thiago Faria and Gimenes، نويسنده , , Denise Tofanello and Richter، نويسنده , , Eduardo Mathias and Munoz، نويسنده , , Rodrigo Alejandro Abarza and Lago، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2011
Pages
5
From page
1274
To page
1278
Abstract
We report here, for the first time, application of batch injection analysis (BIA) with amperometric detection for determination of the phenolic antioxidant butylated hydroxyanisole (BHA) in biodiesel. A sample plug was directly injected onto a boron-doped diamond electrode immersed in 50% v/v hydroethanolic solution with 0.1 mol L−1 HClO4 using an electronic micropipette. Importantly, the only preparation step required for biodiesel analysis is dilution in the same hydroethanolic electrolyte solution. Our proposed method has several advantages for routine biodiesel analysis, including: a low relative standard deviation between injections (0.29%, n = 20), high analytical frequency (120 h−1), adequate recovery values (93–101%) for spiked samples, satisfactory accuracy (based on comparative determinations by high-performance liquid-chromatography), and a low detection limit (100 ng of BHA per g of biodiesel). Finally, our method can be adapted for the determination of other antioxidants in biodiesel samples.
Keywords
biodiesel , Amperometry , Boron-doped diamond electrode , antioxidants , Batch injection analysis
Journal title
Talanta
Serial Year
2011
Journal title
Talanta
Record number
1663090
Link To Document