Title of article :
On-line coupling of automatic solid-phase extraction and HPLC for determination of carotenoids in serum
Author/Authors :
Ferreiro-Vera، نويسنده , , C. and Mata-Granados، نويسنده , , J.M. and Quesada Gَmez، نويسنده , , J.M. and Luque de Castro، نويسنده , , M.D.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2011
Abstract :
The automated method developed for the determination of carotenoids uses 200 μL of serum, which was mixed with 400 μL of tetrahydrofuran, vortexed for 1 min, settled for 10 min, centrifuged for 6 min and the supernatant injected into an automatic solid-phase extraction (SPE) system for cleanup–preconcentration. A 10% water–acetonitrile mobile phase at 1.5 mL min−1 eluted the retained compounds and transferred them on-line to a reversed-phase analytical column for individual separation of the target analytes. Visible detection was performed at 450 and 460 nm. The detection limits for the target analytes were between 3 and 30 ng mL−1; the precision (expressed as relative standard deviation) ranged between 2.83 and 5.06% for repeatability and between 3.80 and 7.40% for within laboratory reproducibility. The total analysis time was 18 min. The proposed method is reliable, robust, and has an excellent potential for high-throughput use in both clinical and research laboratories.
Keywords :
serum , On-line SPE , HPLC–DAD , carotenoids