    Selectivity coefficients of ion-selective magnesium electrodes used for simultaneous determination of magnesium and calcium ions

  • Author/Authors

    Maj-Zurawska، نويسنده , , Magdalena and Lewenstam، نويسنده , , Andrzej، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2011
  • Pages
    7
  • From page
    295
  • To page
    301
  • Abstract
    Membrane ion-selective magnesium electrodes are commonly used to determine ionized magnesium concentration in blood serum and intracellular fluid by potentiometric clinical analyzers. The selectivity of these electrodes against calcium ion is typically insufficient to avoid calcium interference in blood serum analysis. For this reason the selectivity coefficient for calcium ion has to be studied to make possible any mathematical corrections for calcium ion influence. Existing methods relate to the thermodynamic concept of ISE response which suggest a single constant value of the selectivity coefficient and slope that are stable over the concentration ranges of calcium and magnesium ions in the samples. Unfortunately, this rarely happens, and we rather observe dependences on solution and membrane composition, readout time, matrices (anticoagulant, vial coats) that justify usage of apparent selectivities and slopes. To get the practical insight into the response of magnesium ion-selective electrodes a novel method for estimating the selectivity coefficients and the slope of the electrode characteristics is proposed. This method is an effective starting point for selecting electrodes and designing transient signal software in a potentiometric clinical analyzer. The method allows obtaining the ionized magnesium concentration in blood serum with minimal possible error by addressing the assessed targets, i.e. apparent selectivity and slope. The method is based on computer simulation and on the Nicolsky–Eisenman equation. Usually only a few iterations are needed to obtain stable congruent results. The method presented is particularly useful in conditions where is not possible to obtain calibration curve, which is typical for clinical analyzer where at most three point calibration is performed.
  • Keywords
    Selectivity coefficients , Ion-selective magnesium electrode , ionized magnesium , Clinical analysis
  • Journal title
    Talanta
  • Serial Year
    2011
  • Journal title
    Talanta
  • Record number

    1663909