Title of article :
Design and evaluation of an electrochemical immunosensor for measles serodiagnosis using measles-specific Immunoglobulin G antibodies
Author/Authors :
Mashazi، نويسنده , , Philani and Vilakazi، نويسنده , , Sibulelo and Nyokong، نويسنده , , Tebello، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2013
Abstract :
The design of electrochemical immunosensors for the detection of measles-specific antibodies is reported. The measles-antigen modified surface was used as an antibody capture surface. The detection of measles-specific IgG antibodies was accomplished using the voltammetric method and horse-radish peroxidase (HRP) labeled secondary antibody (anti-IgG) as a detecting antibody. The potential applications of the designed immunosensor were evaluated in buffer and serum solutions. The immunosensor exhibited good linearity at concentrations less than 100 ng mL−1 with R2=0.997 and the limit of detection of 6.60 ng mL−1 at 3σ. The potential application of the immunosensor was evaluated in the deliberately infected human and newborn calf serum samples with measles-IgG antibody mimicking real-life samples. The designed electrochemical immunosensor could differentiate between infected and un-infected serum samples as higher catalytic currents were obtained for infected serum samples.
Keywords :
Immunosensors , Measles-specific antibodies , Horse-radish peroxidase , Labeled-detection , Measles-antigen