Title of article :
A simple recirculating flow system for the calibration of polar organic chemical integrative samplers (POCIS): Effect of flow rate on different water pollutants
Author/Authors :
Di Carro، نويسنده , , Marina and Bono، نويسنده , , Luca and Magi، نويسنده , , Emanuele، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2014
Pages :
4
From page :
30
To page :
33
Abstract :
A calibration system for POCIS was developed and used to calculate the sampling rates of eight analytes belonging to pesticides, non-steroidal anti-inflammatory drugs and perfluorinated compounds: atrazine, propazine, terbutylazine, diclofenac, ibuprofen, ketoprofen, perfluorooctanoic acid and perfluorooctanesulfonate. Experiments with a linear velocity of 2.0, 5.1, 10.2 and 15.3 cm/s were carried out for 96 h using two different analyte concentrations. POCIS extracts were analyzed by liquid chromatography–tandem mass spectrometry (LC–MS/MS), using multiple reaction monitoring to maximize sensitivity. Results highlighted that the calculated sampling rates are rather constant at the considered concentrations and flow rates. Obtained values of sampling rates were then employed to calculate Time-Weighted Average concentration of the analytes in river and drinking waters.
Keywords :
sampling rate , Time-Weighted Average concentration (TWA) , Polar organic chemical integrative sampler (POCIS) , Polar organic micropollutant , Water monitoring
Journal title :
Talanta
Serial Year :
2014
Journal title :
Talanta
Record number :
1670055
Link To Document :
بازگشت