• Title of article

    A bimodal optoelectronic flow-through detector for phosphate determination

  • Author/Authors

    Marta and Fiedoruk، نويسنده , , Marta and Mieczkowska، نويسنده , , El?bieta and Koncki، نويسنده , , Robert and Tymecki، نويسنده , , ?ukasz، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2014
  • Pages
    4
  • From page
    211
  • To page
    214
  • Abstract
    A miniature flow-through detector useful for bimodal, photometric and fluorimetric, determination of phosphates has been developed. This optoelectronic device made of four light emitting diodes (LEDs) integrated in the form of 85 µL optical cell is easily applied in flow analysis manifolds. These LEDs play the roles of light source for photometric measurements, fluorescence inductors and detector of absorbance and fluorescence. For photometric mode of determinations a phosphomolybdenum blue method has been applied. The fluorimetric method of phosphate determination is based on quenching of rhodamine fluorescence by the heteropolyacid. The developed detector used in a simple three-channel flow injection analysis (FIA) system allows photometric or fluorimetric determination of phosphate in the wide range of concentration. The detection limits found for photometric and fluorimetric modes of FIA measurements are 5.5 mg L−1 and 10.4 µg L−1, respectively. The potential utility of the flow-through detector for the needs of food and clinical analysis has been demonstrated.
  • Keywords
    Optoelectronic detector , Fluorimetry , Phosphate , Photometry , Flow Analysis
  • Journal title
    Talanta
  • Serial Year
    2014
  • Journal title
    Talanta
  • Record number

    1671332