Title of article :
Sensitive electrochemical detection of dopamine with a DNA/graphene bi-layer modified carbon ionic liquid electrode
Author/Authors :
Wang، نويسنده , , Xiaofeng and You، نويسنده , , Zheng and Sha، نويسنده , , Hailiang and Cheng، نويسنده , , Yong and Zhu، نويسنده , , Huanhuan and Sun، نويسنده , , Wei، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2014
Pages :
6
From page :
373
To page :
378
Abstract :
A DNA and graphene (GR) bi-layer modified carbon ionic liquid electrode (CILE) was fabricated by an electrodeposition method. GR nanosheets were electrodeposited on the surface of CILE at the potential of −1.3 V and then DNA was further deposited at the potential of +0.5 V on GR modified CILE. Electrochemical performances of the fabricated DNA/GR/CILE were carefully investigated. Then electrochemical behaviors of dopamine (DA) on the modified electrode were studied with the calculated electrochemical parameters. Under the optimized conditions, a linear relationship between the oxidation peak current and the concentration of DA was obtained in the range from 0.1 μmol/L to 1.0 mmol/L with a detection limit of 0.027 μmol/L (3σ). The modified electrode exhibited excellent reproducibility, repeatability, stability, validation and robustness for the electrochemical detection of DA. The proposed method was further applied to the DA injection solution and human urine samples determination with satisfactory results.
Keywords :
Grapheme , dsDNA , Dopamine , Carbon ionic liquid electrode , Differential pulse voltammetry