Title of article :
Multielemental characterisation of cobalt by glow discharge quadrupole mass spectrometry
Author/Authors :
Raparthi، نويسنده , , Shekhar and Arunachalam، نويسنده , , J. and Das، نويسنده , , Niranjan and Srirama Murthy، نويسنده , , A.M.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2005
Pages :
9
From page :
1270
To page :
1278
Abstract :
Multielemental determination and the assessment of purity of cobalt metal used in the preparation of Ni-based super-alloys have been carried out by glow discharge quadrupole mass spectrometry (GD-QMS). Relative sensitivity factors (RSF) generated from certified iron matrix reference samples (NIST 663 and 664 low alloy steel pin standards) could be used for the determination of different trace element constituents of the sample. Different wet chemical procedures were also carried out for the determination of the trace constituents in the sample. The GD-QMS results are in reasonably good agreement with those obtained from wet chemical procedures, validating the use of the RSF values generated on low alloy steel standards for the computation of trace element concentrations in cobalt metal. A variety of molecular ions formed through the reaction of cobalt (matrix) with the discharge gas (argon) were also detected.
Keywords :
Glow discharge quadrupole mass spectrometry , Multielemental characterisation , Cobalt , Relative sensitivity factors
Journal title :
Talanta
Serial Year :
2005
Journal title :
Talanta
Record number :
1674159