• Title of article

    Evaluation of a high sensitivity PbO2 pH-sensor

  • Author/Authors

    Lima، نويسنده , , Alexandre Correa and Jesus، نويسنده , , Adriana Aparecida and Tenan، نويسنده , , Mلrio Alberto and Silva، نويسنده , , Astréa F. de Souza and Oliveira، نويسنده , , André Fernando and Fatibello-Filho، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2005
  • Pages
    4
  • From page
    225
  • To page
    228
  • Abstract
    In this paper, a new high sensitivity potentiometric pH-electrode with a response in the acid region is proposed. It consists of a PbO2–paraffin matrix deposited on graphite. Its simple construction and studies of some variables at 0.5 mol L−1 ionic strength are reported. A direct relationship between the electric potential difference and the solution pH was observed for pH values ranging from 1.2 to 7.5. A slope greater than 100 mV/decade and a conditional electric potential of about 1250 mV were obtained. The results presented a high correlation with those from a conventional glass pH-electrode in complex matrix samples.
  • Keywords
    High sensitivity pH-electrode , pH-sensor , PbO2–paraffin membrane
  • Journal title
    Talanta
  • Serial Year
    2005
  • Journal title
    Talanta
  • Record number