Title of article :
A highly selective thallium(I) electrode based on a thia substituted macrocyclic ionophore
Author/Authors :
Singh، نويسنده , , Ashok Kumar and Saxena، نويسنده , , Puja، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2005
Pages :
6
From page :
993
To page :
998
Abstract :
A new highly Tl(I)-selective PVC membrane electrode based on tetrathia macrocycle 6,7: 14,15-dibenzo-5,8,13,16-tetraoxo-1,4,9,12-tetrathiacyclohexadecane [Bz2O4(16)aneS4] (I) as membrane carrier, o-nitrophenyloctyl ether (o-NPOE) as solvent mediator and potassium tetrakis(p-chlorophenyl)borate (KTpClPB) as lipophilic additive has been developed. The best performance was given by the membrane of macrocycle (I) with composition 3:120:1.5:50 (I:o-NPOE:KTpClPB:PVC). This electrode exhibits a Nernstian response to Tl(I) ions in the concentration range 1.0 × 10−1–2.23 × 10−6 M with a slope of 58.2 mV/decade of concentration and a detection limit of 1.58 × 10−6 M. The response time of the sensor is 12 s and can be used over a period of 4 months with good reproducibility. The proposed electrode revealed good selectivity over a wide variety of other cations including alkali, alkaline earth, heavy and transition metals. The electrode works well over a pH range of 3.2–11.5 and in partially non-aqueous medium with up to 30% organic content. The sensor was also used as an indicator electrode in potentiometric titration of Tl(I) ions with KI solution.
Keywords :
Tl(I)-selective electrode , Tetrathia macrocycle , PVC-based membrane , Potentiometry
Journal title :
Talanta
Serial Year :
2005
Journal title :
Talanta
Record number :
1674369
Link To Document :
بازگشت