    A peroxidase-tetrathiafulvalene biosensor based on self-assembled monolayer modified Au electrodes for the flow-injection determination of hydrogen peroxide

  • Author/Authors

    Campuzano، نويسنده , , Susana and Pedrero، نويسنده , , Marيa and Pingarrَn، نويسنده , , José M.، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2005
  • Pages
    10
  • From page
    1310
  • To page
    1319
  • Abstract
    The construction and performance under flow-injection conditions of an integrated amperometric biosensor for hydrogen peroxide is reported. The design of the bioelectrode is based on a mercaptopropionic acid (MPA) self-assembled monolayer (SAM) modified gold disk electrode on which horseradish peroxidase (HRP, 24.3 U) was immobilized by cross-linking with glutaraldehyde together with the mediator tetrathiafulvalene (TTF, 1 μmol), which was entrapped in the three-dimensional aggregate formed. perometric biosensor allows the obtention of reproducible flow injection amperometric responses at an applied potential of 0.00 V in 0.05 mol L−1 phosphate buffer, pH 7.0 (flow rate: 1.40 mL min−1, injection volume: 150 μL), with a range of linearity for hydrogen peroxide within the 2.0 × 10−7–1.0 × 10−4 mol L−1 concentration range (slope: (2.33 ± 0.02) × 10−2 A mol−1 L, r = 0.999). A detection limit of 6.9 × 10−8 mol L−1 was obtained together with a R.S.D. (n = 50) of 2.7% for a hydrogen peroxide concentration level of 5.0 × 10−5 mol L−1. The immobilization method showed a good reproducibility with a R.S.D. of 5.3% for five different electrodes. Moreover, the useful lifetime of one single biosensor was estimated in 13 days. M-based biosensor was applied for the determination of hydrogen peroxide in rainwater and in a hair dye. The results obtained were validated by comparison with those obtained with a spectrophotometric reference method. In addition, the recovery of hydrogen peroxide in sterilised milk was tested.
  • Keywords
    Hydrogen peroxide , Self-assembled monolayers , tetrathiafulvalene , Amperometric biosensor
  • Journal title
    Talanta
  • Serial Year
    2005
  • Journal title
    Talanta
  • Record number

    1674441