• Title of article

    A spectrophotometric procedure for DNA assay with a micro-sequential injection lab-on-valve meso-fluidic system

  • Author/Authors

    Chen، نويسنده , , Xuwei and Wang، نويسنده , , Jianhua and Fang، نويسنده , , Zhaolun، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2005
  • Pages
    6
  • From page
    227
  • To page
    232
  • Abstract
    A micro-sequential injection spectrophotometric procedure for DNA assay was developed based on the employment of a lab-on-valve (LOV) meso-fluidic analytical system. A small amount of crystal violet solution (10 μl) was de-colored inside the flow cell of the LOV at the presence of 5 μl λ-DNA/HindIII within a certain pH range, and the absorbance decrease of crystal violet solution at 591 nm was measured via optical fibers and was employed as the basis of quantification. A uni-variant approach was adopted for the optimization of experimental parameters, including buffer pH, concentration and volume of crystal violet solution, reaction time and sample/reagent loading flow rates. A linear calibration graph was obtained within 0.2–6.0 μg ml−1, along with a detection limit of 0.07 μg ml−1. The procedure was applied for the determination of λ-DNA/HindIII in synthetic samples in comparison with a documented procedure.
  • Keywords
    Sequential injection , Lab-on-valve , Meso-fluidic analytical system , DNA assay , Spectrophotometry
  • Journal title
    Talanta
  • Serial Year
    2005
  • Journal title
    Talanta
  • Record number

    1674503