• Title of article

    On-line determination of trace sulfur dioxide in air by integrated microchip coupled with fluorescence detection

  • Author/Authors

    Chang، نويسنده , , Wei and Ono، نويسنده , , Yasunari and Kumemura، نويسنده , , Momoko and Korenaga، نويسنده , , Takashi، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2005
  • Pages
    5
  • From page
    646
  • To page
    650
  • Abstract
    A microchip-based method was developed for on-line determination of trace sulfur dioxide (SO2) in air. Gaseous SO2, which diffused through the porous glass materials on the microchip, was absorbed into an absorption solution of triethanolamine (TEA) as sulfite ions and reacted with N-(9-acridinyl)maleimide (NAM), which was used as a fluorescent reagent. The fluorescence of NAM-sulfite in micro-fluidic channel was detected. The calibration curve of sulfite ions in the range of 1.5–30 μmol/L (SO2 3–60 ppbv) showed a linear relation R2 = 0.995, and the relative standard deviation (R.S.D.) was 1.9% for 10 μmol/L sulfite ions in five measurements. The entire measurement procedure was achieved by the integrated microchip, and the consumption of reagents was drastically reduced. It was satisfactory to apply this method to determine on-line the SO2 level in the air.
  • Keywords
    Microchip , fluorescence detection , Sulfur dioxide , On-line monitoring device
  • Journal title
    Talanta
  • Serial Year
    2005
  • Journal title
    Talanta
  • Record number