Title of article :
Differential cyclic voltammetry for selective and amplified detection
Author/Authors :
van Megen، نويسنده , , M.J.J. and Odijk، نويسنده , , M. and Wiedemair، نويسنده , , J. and Olthuis، نويسنده , , W. and van den Berg، نويسنده , , A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Pages :
5
From page :
6
To page :
10
Abstract :
We propose to combine two existing electrochemical techniques, cyclic voltammetry (CV) and redox cycling (RC), in order to obtain amplified and selective detection of redox active species. This combination is achieved by applying CV waveforms to two electrodes spaced 1.20 μm apart, with one of the electrodes at a constant potential offset compared to the other. Due to this potential offset we have named this technique differential cyclic voltammetry (DCV). Analytical expressions for the DCV voltammogram are derived and an optimal potential offset is calculated. The optimal voltage difference for the trade-off between peak height and width is 0.1 V for redox couples with n = 1. Experimental voltammograms show good agreement with the analytical expressions. The voltammogram for ferrocenedimethanol has been fitted (R2 = 0.9985) using only the distance between the electrodes as fitting parameter. Therefore, this technique shows promise as a tool for amplified and selective detection of redox active species.
Keywords :
Redox cycling , SECM , Thin layer cells , Cyclic voltammetry
Journal title :
Journal of Electroanalytical Chemistry
Serial Year :
2012
Journal title :
Journal of Electroanalytical Chemistry
Record number :
1676026
Link To Document :
بازگشت