Title of article :
Vanadium determination in water using alkaline phosphatase based screen-printed carbon electrodes modified with gold nanoparticles
Author/Authors :
Alvarado-Gلmez، نويسنده , , Ana L. and Alonso-Lomillo، نويسنده , , M.A. and Domيnguez-Renedo، نويسنده , , O. and Arcos-Martيnez، نويسنده , , M.J.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Pages :
5
From page :
51
To page :
55
Abstract :
A chronoamperometric assay for vanadium ions, based on the inhibition of the enzyme alkaline phosphatase, is reported. Screen-printed carbon electrodes modified with gold nanoparticles were used as transducers for the immobilization of the enzyme. The enzymatic activity over riboflavin-5-monophosphate sodium salt is affected by vanadium ions, which results in a decrease in the chronoamperometric current registered. The assay has a detection limit of 3.9 ± 0.4 μM, a repeatability of 7.6% (n = 4) and a reproducibility of 11.6% (n = 4). A study of the possible interferences shows that the presence of Mo (VI) may affect vanadium determination. The method was successfully applied to the determination of vanadium in tap water.
Keywords :
Biosensor , Riboflavin 5-monophosphate , alkaline phosphatase , vanadium , Screen printed carbon electrode
Journal title :
Journal of Electroanalytical Chemistry
Serial Year :
2013
Journal title :
Journal of Electroanalytical Chemistry