Title of article :
On-line preconcentration of cobalt in drinking water using a minicolumn packed with activated carbon coupled to electrothermal atomic absorption spectrometric determination
Author/Authors :
Cerutti، نويسنده , , Soledad and Moyano، نويسنده , , Susana and Gلsquez، نويسنده , , José A. and Stripeikis، نويسنده , , Jorge and Olsina، نويسنده , , Roberto A. and Martinez، نويسنده , , Luis D.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
7
From page :
2015
To page :
2021
Abstract :
An on-line flow injection preconcentration-electrothermal atomic absorption spectrometry method is developed for trace determination of cobalt in drinking water samples by sorption on a conical minicolumn packed with activated carbon at pH 9.5. The cobalt was eluted from the minicolumn with 10% (v/v) nitric acid. An enrichment factor of 190-fold for a sample volume of 10 ml was obtained. The detection limit (DL) value for the preconcentration method proposed was 5 ng l−1. The precision for 10 replicate determinations at the 50 ng l−1 Co level was 4.7% relative standard deviation. The calibration graph using the preconcentration system for cobalt was linear with a correlation coefficient of 0.9993 at levels near the DLs up to at least 0.35 μg l−1. The method was successfully applied to the determination of cobalt in drinking water samples.
Keywords :
Cobalt determination , On-line preconcentration , Electrothermal atomic absorption spectrometry , Activated carbonwater samples
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Serial Year :
2003
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Record number :
1679930
Link To Document :
بازگشت