Title of article :
Comparison of conventional and total reflection excitation geometry for fluorescence X-ray absorption spectroscopy on droplet samples
Author/Authors :
Falkenberg، نويسنده , , G. and Pepponi، نويسنده , , G. and Streli، نويسنده , , C. and Wobrauschek، نويسنده , , P.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
6
From page :
2239
To page :
2244
Abstract :
X-ray absorption fine structure (XAFS) experiments in fluorescence mode have been performed in total reflection excitation geometry and conventional 45°/45° excitation/detection geometry for comparison. The experimental results have shown that XAFS measurements are feasible under normal total reflection X-ray fluorescence (TXRF) conditions, i.e. on droplet samples, with excitation in grazing incidence and using a TXRF experimental chamber. The application of the total reflection excitation geometry for XAFS measurements increases the sensitivity compared to the conventional geometry leading to lower accessible concentration ranges. However, XAFS under total reflection excitation condition fails for highly concentrated samples because of the self-absorption effect.
Keywords :
X-ray Absorption Fine Structure Spectroscopy , Total reflection , Fluorescence mode , Self-absorption , pb
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Serial Year :
2003
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Record number :
1680030
Link To Document :
بازگشت