• Title of article

    A hydride generation flow system for determination of arsenic and selenium by total reflection X-ray fluorescence spectrometry

  • Author/Authors

    Gomes، نويسنده , , Ana Carla F. and Meneg?rio، نويسنده , , Amauri A. and Pellegrinotti، نويسنده , , Daniel C. and Giné، نويسنده , , Maria F. and Nascimento Filho، نويسنده , , Virg??lio F.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    4
  • From page
    1481
  • To page
    1484
  • Abstract
    In this work, a preconcentration and separation system based on continuous flow hydride generation is proposed to improve the determination of As and Se by total reflection X-ray fluorescence spectrometry. The generated hydrides are continuously separated from the liquid phase and collected in a chamber containing 250 μl of HCl/HNO3 1:1 (v/v) solution. Hydride generation conditions and collection of the hydrides were evaluated. Under optimised conditions, enrichment factors of 55 for As and 82 for Se were attained. Detection limits of 0.3 μg l−1 for As and Se were obtained when 20 ml of sample was used. Analysis of a natural water standard reference material from National Institute of Standard and Technology (SRM-1640) was in agreement with the certified values at the 95% confidence level.
  • Keywords
    Hydride generation , TXRF , Preconcentration , Flow system
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Serial Year
    2004
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Record number