• Title of article

    Measurement of gas kinetic temperatures for polyatomic ions in inductively coupled plasma-mass spectrometry: Validation and refinements

  • Author/Authors

    McIntyre، نويسنده , , Sally M. and Ferguson، نويسنده , , Jill Wisnewski and Witte، نويسنده , , Travis M. and Houk، نويسنده , , R.S.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    7
  • From page
    248
  • To page
    254
  • Abstract
    The general method of comparing measured ion ratios to calculated ion ratios to determine a gas kinetic temperature (Tgas) is reviewed. Various mathematical refinements to the calculated partition functions are examined for their effect on the determined Tgas. It is found that (a) excited electronic states should be included for ArO+, neutral NO, and O2; (b) a 10% error in solvent load, sample gas flow rate, vibrational constant (ω), rotational constant (B) or measured ion ratio produces only a 1 to 3% error in Tgas; (c) a 10% error in dissociation energy (D0) creates nearly a 10% error in Tgas; and (d) high temperature corrections to the partition functions produce minimal change and can generally be neglected.
  • Keywords
    Ion extraction ICP-MS , inductively coupled plasma mass spectrometry , Inductively coupled plasma , ICP-MS
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Serial Year
    2011
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Record number