Title of article
Measurement of gas kinetic temperatures for polyatomic ions in inductively coupled plasma-mass spectrometry: Validation and refinements
Author/Authors
McIntyre، نويسنده , , Sally M. and Ferguson، نويسنده , , Jill Wisnewski and Witte، نويسنده , , Travis M. and Houk، نويسنده , , R.S.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2011
Pages
7
From page
248
To page
254
Abstract
The general method of comparing measured ion ratios to calculated ion ratios to determine a gas kinetic temperature (Tgas) is reviewed. Various mathematical refinements to the calculated partition functions are examined for their effect on the determined Tgas. It is found that (a) excited electronic states should be included for ArO+, neutral NO, and O2; (b) a 10% error in solvent load, sample gas flow rate, vibrational constant (ω), rotational constant (B) or measured ion ratio produces only a 1 to 3% error in Tgas; (c) a 10% error in dissociation energy (D0) creates nearly a 10% error in Tgas; and (d) high temperature corrections to the partition functions produce minimal change and can generally be neglected.
Keywords
Ion extraction ICP-MS , inductively coupled plasma mass spectrometry , Inductively coupled plasma , ICP-MS
Journal title
Spectrochimica Acta Part B Atomic Spectroscopy
Serial Year
2011
Journal title
Spectrochimica Acta Part B Atomic Spectroscopy
Record number
1681046
Link To Document