Title of article
Calibration of wavelength-dispersive X-ray spectrometer for standardless analysis
Author/Authors
Sitko، نويسنده , , Rafal and Zawisza، نويسنده , , Beata، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2005
Pages
6
From page
95
To page
100
Abstract
An experimental procedure for calibrating wavelength-dispersive X-ray spectrometer is presented. The detection efficiency was determined from the ratio of fluorescent radiation measured using the flow-proportional and the scintillation counter. The relative reflection efficiency of the analyzing crystal for various wavelengths was determined from the X-ray measurements of standard samples. The absorption correction for the atmosphere in the spectrometer chamber was also experimentally examined. All parameters (with the exception of reflection efficiency of the analyzing crystal) can be calculated from the X-ray measurements of unknown samples.
Keywords
Calibration , WDXRF , Reflection efficiency , fundamental parameters , Detection efficiency
Journal title
Spectrochimica Acta Part B Atomic Spectroscopy
Serial Year
2005
Journal title
Spectrochimica Acta Part B Atomic Spectroscopy
Record number
1681314
Link To Document