Matrix effect on the detection limit and accuracy in total reflection X-ray fluorescence analysis of trace elements in environmental and biological samples
Author/Authors
J. Karjou، نويسنده , , J.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
5
From page
177
To page
181
Abstract
The effect of matrix contents on the detection limit of total reflection X-ray fluorescence analysis was experimentally investigated using a set of multielement standard solutions (500 ng/mL of each element) in variable concentrations of NH4NO3. It was found that high matrix concentration, i.e. 0.1–10% NH4NO3, had a strong effect on the detection limits for all investigated elements, whereas no effect was observed at lower matrix concentration, i.e. 0–0.1% NH4NO3.
fect of soil and blood sample masses on the detection limit was also studied. The results showed decreasing the detection limit (in concentration unit, μg/g) with increasing the sample mass. However, the detection limit increased (in mass unit, ng) with increasing sample mass. The optimal blood sample mass of ca. 200 μg was sufficient to improve the detection limit of Se determination by total reflection X-ray fluorescence.
pability of total reflection X-ray fluorescence to analyze different kinds of samples was discussed with respect to the accuracy and detection limits based on certified and reference materials. Direct analysis of unknown water samples from several sources was also presented in this work.
Keywords
detection limit , Direct analysis , TXRF , reference material
Journal title
Spectrochimica Acta Part B Atomic Spectroscopy