Title of article :
A new approach for fluorine determination by solid sampling graphite furnace molecular absorption spectrometry
Author/Authors :
de Moraes Flores، نويسنده , , ةder Lisandro and Barin، نويسنده , , Juliano Smanioto and de Moraes Flores، نويسنده , , ةrico Marlon and Dressler، نويسنده , , Valderi Luiz، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
This work deals with the determination of fluorine by solid sampling graphite furnace molecular absorption spectrometry. The molecular absorbance of aluminum monofluoride (AlF), which is produced in the vapor phase in the presence of Al3+, is measured at 227.5 nm, a non-resonant platinum line. A conventional graphite furnace program has been used with pyrolysis and vaporization temperatures of 800 and 2300 °C, respectively. Solutions of Ba2+ and Al3+ have been used to avoid fluorine losses during the pyrolysis stage and to produce AlF in the vaporization stage, respectively. Certified coal and alumina samples were analyzed using aqueous standards for calibration. The agreement between the found concentration and the certified value, or the value obtained by another method ranged from 92 to 105%, with a relative standard deviation less than 8.5%. The limit of detection and the characteristic mass was 0.17 μg g− 1 and 205 pg, respectively.
Keywords :
alumina , Solid sampling , Graphite furnace molecular absorption spectrometry , Coal , fluorine
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy