Title of article :
Compact fiber-optic fluorosensor employing light-emitting ultraviolet diodes as excitation sources
Author/Authors :
Ek، نويسنده , , Sara and Anderson، نويسنده , , Benjamin and Svanberg، نويسنده , , Sune، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Abstract :
A compact fluorosensor using three different ultraviolet light-emission diodes as excitation sources for fiber-optic recording of fluorescence spectra from samples is described. A compact integrated spectrometer with linear array wavelength recording is used, yielding a spectral resolution of about 8 nm. In two system implementations ultraviolet light-emitting diodes at 300, 340 and 395 nm, or at 360, 385 and 410 nm were used as excitation sources with typical emission halfwidths of 12 nm, each combined with a matching long-path colored-glass filter automatically brought into the fluorescence light flow for suppression of reflected light. Spectra from measurements on vegetation, human skin tumors and a rare-earth ion-based thermographic phosphor were recorded to illustrate the system performance.
Keywords :
Spectrometer , Cancer diagnostic , UV light-emitting diode , fluorescence
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy