Title of article
Compact fiber-optic fluorosensor employing light-emitting ultraviolet diodes as excitation sources
Author/Authors
Ek، نويسنده , , Sara and Anderson، نويسنده , , Benjamin and Svanberg، نويسنده , , Sune، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
5
From page
349
To page
353
Abstract
A compact fluorosensor using three different ultraviolet light-emission diodes as excitation sources for fiber-optic recording of fluorescence spectra from samples is described. A compact integrated spectrometer with linear array wavelength recording is used, yielding a spectral resolution of about 8 nm. In two system implementations ultraviolet light-emitting diodes at 300, 340 and 395 nm, or at 360, 385 and 410 nm were used as excitation sources with typical emission halfwidths of 12 nm, each combined with a matching long-path colored-glass filter automatically brought into the fluorescence light flow for suppression of reflected light. Spectra from measurements on vegetation, human skin tumors and a rare-earth ion-based thermographic phosphor were recorded to illustrate the system performance.
Keywords
Spectrometer , Cancer diagnostic , UV light-emitting diode , fluorescence
Journal title
Spectrochimica Acta Part B Atomic Spectroscopy
Serial Year
2008
Journal title
Spectrochimica Acta Part B Atomic Spectroscopy
Record number
1682213
Link To Document