• Title of article

    Laser-induced breakdown spectroscopy—From research to industry, new frontiers for process control

  • Author/Authors

    Noll، نويسنده , , Reinhard and Sturm، نويسنده , , Volker and Aydin، نويسنده , , ـmit and Eilers، نويسنده , , Daniel and Gehlen، نويسنده , , Christoph and Hِhne، نويسنده , , Manuela and Lamott، نويسنده , , André and Makowe، نويسنده , , Joachim and Vrenegor، نويسنده , , Jens، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    8
  • From page
    1159
  • To page
    1166
  • Abstract
    This paper presents R&D activities to explore new laser parameter ranges in pulse energy, time and space for laser-induced breakdown spectroscopy. The collinear double pulse effect, which is well studied for pulses of typically several 100 mJ energy can also be observed for laser pulses having a pulse energy two orders of magnitude lower. In this case, maximum line emission intensity occurs at interpulse separations of a few 100 ns. Temporal pulse tailoring to improve the performance of LIBS is only a first step. A comprehensive approach includes spatial pulse shaping to generate craters with predefined shape or to improve spatial averaging for the analysis of inhomogeneous samples. High performance components for LIBS systems such as spectrometers, electronics and sample stands are required to enable industrial applications. Latest developments offer wide-band single spectra acquisition with a high spectral resolution at a measuring frequency of up to 500 Hz. The next generation of multi-channel integrator electronics for Paschen–Runge spectrometers equipped with PMT detectors will further push the measuring speed to up to 5 kHz, thus opening a new area of high-speed LIBS microanalysis. Novel LIBS devices for various industrial applications presented include analysis of metallic process control samples with scale layers, on-site analysis of slag samples in secondary metallurgy, high-speed identification of Al scrap, mix-up detection of pipe fittings as well as recent work towards in-process identification of hot coils in a rolling mill.
  • Keywords
    LIBS application , LIBS , Spatial pulse tailoring , Sample stand , Temporal pulse tailoring , Spectrometer
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Serial Year
    2008
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Record number

    1682472