Title of article :
Inductively coupled plasma mass spectrometry with ambient helium surrounding ion source
Author/Authors :
Jin، نويسنده , , Xianzhong and Li، نويسنده , , Lingfeng and Hang، نويسنده , , Wei and Chen، نويسنده , , Jianguo and Huang، نويسنده , , Benli، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Pages :
4
From page :
1052
To page :
1055
Abstract :
An enclosed device surrounding the argon inductively coupled plasma torch was fabricated to exclude air entrainment and attenuate background interferences. Helium was introduced into the enclosure, and ambient helium plasma was formed stably. Under cold plasma condition, we found that the spectral background decreased about 1 order of magnitude averagely compared with that in typical operation condition. For laser ablation with a Nd:YAG laser, the limits of detection of 28Si, 29Si, 31P, and 32S in an iron matrix were improved significantly; the linearity of their calibration curves was greatly improved as well compared with standard mode and cool mode ICP-MS with no ambient helium. The result indicates that polyatomic interferences from nitrogen, oxygen, hydrogen, carbon, etc. were effectively reduced in helium ambient ICP-MS.
Keywords :
Laser ablation , Gas species interferences , inductively coupled plasma mass spectrometry , Enclosed plasma device , Ambient helium
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Serial Year :
2010
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Record number :
1683693
Link To Document :
بازگشت