• Title of article

    A technique for estimating contamination tolerance limits for stable isotope ratio determinations

  • Author/Authors

    Doherty، نويسنده , , William، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    8
  • From page
    114
  • To page
    121
  • Abstract
    A model of the interaction between the precision of an isotope ion signal measurement and the accuracy of an isotope ratio determination was developed and used to derive the equations required to calculate the maximum tolerable amount of contamination for stable isotope ratio determinations. Comparison of the calculated tolerance limits and the blank estimated amount of contaminant will establish whether or not a correction for the contribution of the contaminant to the gross signals will be required. The 1000:1 sample-to-contaminant concentration tolerance limit used in stable isotope ratio plasma mass spectrometry will, in some situations, underestimate the contamination error compared to the model calculations. tion of the limit equations required an empirically determined relation between the signal strength and the signalʹs relative standard deviation (signal noise function). A single hyperbolic signal noise function was used to describe the behaviour of isotope ion signals of Ni, Cu, Tl and Pb measured using a plasma source double focusing magnetic sector mass spectrometer. The derivation could be extended to accommodate different signal noise functions.
  • Keywords
    Blank , Plasma mass spectrometry , Isotope ratio , Signal noise , Contamination error
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Serial Year
    2011
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Record number

    1683733