• Title of article

    Ultrasonic nebulization-sample introduction system for quantitative analysis of liquid samples by laser-induced breakdown spectroscopy

  • Author/Authors

    Aras، نويسنده , , Nadir and Ye?iller، نويسنده , , Semira Unal and Ate?، نويسنده , , Dilek Ar?ca and Yalç?n، نويسنده , , ?erife، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2012
  • Pages
    8
  • From page
    87
  • To page
    94
  • Abstract
    In this study, design and optimization studies of a sample introduction system based on ultrasonic nebulization of metal salts in aqueous environment for laser-induced breakdown spectroscopic detection were presented. The system consisted of an ultrasonic nebulizer connected to a tandem heater–condenser–membrane dryer unit that produces sub-micron size aerosols. Results indicate improvements in detection limits for some elements with the use of membrane dryer. Optimization studies were performed by systematical investigation of LIBS emission signal with respect to laser energy, carrier gas flow rate and detector timing parameters. Under optimized conditions, calibration graphs for Na, K, Mg, Ca, Cu, Al, Cr, Cd, Pb and Zn were constructed and detection limits were calculated. The applicability of the ultrasonic nebulization-LIBS system was tested on real water samples. This system establishes LIBS as an effective analytical tool for both qualitative and quantitative determination of metal aerosols in aqueous environments. This technique is sufficiently rapid to provide real-time monitoring of toxic metals.
  • Keywords
    Laser-induced breakdown spectroscopy , UltraSOnic nebulization , Metal aerosol , Membrane dryer
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Serial Year
    2012
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Record number