Title of article
Considerations about the detection efficiency in inductively coupled plasma mass spectrometry
Author/Authors
Niemax، نويسنده , , Kay، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2012
Pages
5
From page
65
To page
69
Abstract
Experimental investigations of analyte atomization, ionization and diffusion processes in the inductively coupled plasma applying single droplet introduction and optical emission spectroscopy provide hints how to improve the detection efficiency of inductively coupled plasma mass spectrometry. It is discussed how the flow, amount and type of injector gas, the size of droplets injected, the analyte mass, and the sampler interface of the mass spectrometer determine the position of analyte atomization and ionization as well as the magnitude of radial analyte ion diffusion at the interface of the mass spectrometer applied.
Keywords
Microdroplets , ICP-MS , Detection efficiency
Journal title
Spectrochimica Acta Part B Atomic Spectroscopy
Serial Year
2012
Journal title
Spectrochimica Acta Part B Atomic Spectroscopy
Record number
1684339