Title of article :
Double-pulse laser ablation inductively coupled plasma mass spectrometry
Author/Authors :
Gonzلlez، نويسنده , , Jhanis and Liu، نويسنده , , Chunyi and Yoo، نويسنده , , Jong and Mao، نويسنده , , Xianglei and Russo، نويسنده , , Richard E.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
5
From page :
27
To page :
31
Abstract :
This paper describes the use of double-pulse laser ablation to improve ICP-MS internal (temporal relative standard deviation, %TRSD) and external (%RSD) precision. The first laser pulse is used to ablate a large quantity of mass from the sample surface. The second pulse is applied with a variable time delay after the first pulse to break the ablated mass into a finer aerosol, which is more readily transported to and digested in the ICP-MS. A factor of two improvement in %TRSD and factor of five in %RSD are demonstrated.
Keywords :
Double pulse , Laser ablation , ICP-MS , particles , Temporal relative standard deviation , Precision
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Serial Year :
2005
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Record number :
1686555
Link To Document :
بازگشت